Climatic Chambers

    << Back Next >>

 
 
                                  Specification For Climatic Chambers
     
1. Work - Space

: 24 "(L)x 24"(D) x 24"(H)

2. Temperature range : - 20o C to 100o C
3. Temperature tolerance : + / - 1 o C
4. Humidity Range : 95% +/- 3% temp. range 25o C to 60o C
5. Temperature control By solid state digital temperature Indicator Controller for dry and wet bulb with PT-100
5. Power 440 V + / - 5%, 3 phase neutral 50 Hz at 15 amp/phase stabilized AC supply
6. Construction Double wall type with
 
  Interior
  Exterior
  Insulation
  Door
  Gasket
  Hinges and Latches
  Mounting
: 22 swg Stainless Steel
: 20 Swg CRCA MS sheets with powder coating

: 1 00 mm glass wool insulation

: Full front opening door double walled insulated     and interior lined with stainless steel

: Double gasket one on chamber and one on the door

: Heavy duty hinges with locking system

: Trolley mounting type with castors
7. Overall dimesnsions : 32 " (L) x 36" (D) x 60 " (H)
 
  Systems Provided
   
 

• Air-Circulation system

  1/8 Hp capacity continuous rated 230 Volts, single phase AC operated motor along with 7 " dia impeller has been provided for uniform temperature and humidity condition inside the chamber.

• Humidity system

 

Low surface loading inconel tube heaters having capacity 500 watts-2nos. Has been provided to add heat inside the chamber. All the heaters are mounted in the conditioning space and will operate in the black zone.

• Humidity system

 

Low pressure moisture generator tank has been provided to generate the moisture and add humidity in the chamber. 2 KW capacity immersion heater having 1 V4 BSP threading has been provided inside the moisture generator tank. The moisture generator tank has been provided along with low water level controller with mechanical float type refilling arrangement .

• Control panel and wiring

  Separate control panel provided on the top of the chamber which I ncorporate all the indicating and monitoring instruments with control switches, MCB"S , contactors, indicating lamps etc. All the wiring will be done as per IS color code with channel type wiring.

• Utilities provided

  a. Chamber drain
b. Viewing window 30x30 cms 2 pane glass with low wattage heater and manually operated wiper.
c. Chamber interior light.
d. Port hole 25mm dia 1 no.
e. Terminal Panel - 6 nos. 5 amps capacity
f. 30 Lts. capacity S/S tank provided on the top for water storage.

• Water requirement

  30 Lts/day dematerialized water during the humidity test.


Available in customized work-space sizes.                                                

 



Designed by : MID
 Hosted by : PBD  Promoted by : GID


S. No. 16/6, Nanded Phata, Near Savali Dhaba, Pune Sinhagad Road, Pune - 411041
Tel. No. : 0+91-20-